CENELEC - EN 60731
Medical electrical equipment - Dosimeters with ionization chambers as used in radiotherapy
| Organization: | CENELEC |
| Publication Date: | 1 April 2012 |
| Status: | active |
| Page Count: | 108 |
| ICS Code (Radiographic equipment): | 11.040.50 |
scope:
Scope and object
Scope
This International Standard specifies the performance requirements of RADIOTHERAPY DOSIMETERS, intended for the measurement of ABSORBED DOSE TO WATER or AIR KERMA (and their rates and spatial distributions) in PHOTON, ELECTRON, proton or heavy ion RADIATION FIELDS as used in RADIOTHERAPY.
The DOSE MONITORING SYSTEMS incorporated in RADIOTHERAPY treatment machines are not covered by this standard, neither are the re-entrant IONIZATION CHAMBERS used for BRACHYTHERAPY source calibration and constancy check devices.
This standard is applicable to the following types of dosimeter:
a) FIELD-CLASS DOSIMETERS normally used for
1) the measurement of KERMA or dose in a RADIATION BEAM, either in air or in a PHANTOM;
2) in vivo skin surface or intracavitary measurements of dose on PATIENTS.
b) REFERENCE-CLASS DOSIMETERS normally used for the calibration of FIELD-CLASS DOSIMETERS;
NOTE REFERENCE-CLASS DOSIMETERS may be used as FIELD-CLASS DOSIMETERS.
c) SCANNING-CLASS DOSIMETERS normally used for relative dose distribution measurements with a SCANNING SYSTEM such as an automatic water PHANTOM.
Object
The object of this standard is:
- to establish requirements for a satisfactory level of performance for RADIOTHERAPY DOSIMETERS;
- to standardize methods for the determination of compliance with this level of performance.
Three levels of performance are specified:
- a lower level of performance applying to FIELD-CLASS DOSIMETERS;
- a higher level of performance applying to REFERENCE-CLASS DOSIMETERS;
- a specific level of performance applying to SCANNING-CLASS DOSIMETERS.