SNZ AS/NZS 3580.13.2
Methods for sampling and analysis of ambient air Method 13.2: Determination of gaseous and acid-soluble particulate fluorides-Manual, double filter paper sampling
| Organization: | SNZ |
| Publication Date: | 31 May 2013 |
| Status: | active |
| Page Count: | 16 |
scope:
This Standard sets out a method for the collection and separation of gaseous and particulate fluorides from ambient air and their determination using a fluoride ion-selective electrode. Other analytical methods may be used; however, they are outside the scope of this Standard.
This method is applicable to the measurement of gaseous and acid-soluble particulate fluorides suspended in ambient air at concentrations of 0.020 μg/m3 and greater.
For a 24 h sampling period at a flow rate of 0.83 L/s (approximately 50 L/min), an atmospheric concentration of 0.15 μg/m3 of hydrogen fluoride in air can be measured, with measurement uncertainty of 10%.
The lower limit of detection can be improved by sampling for longer periods, to a maximum of seven days (0.020 μg/m3); however, the accumulated particulate matter may interfere with the collection of gaseous fluorides, especially if alumina is present in the particulate matter. The upper limits of detection for gaseous fluoride are 45 μg/m3 for a 24 h sampling period and 6.5 μg/m3 for a seven day sampling period (at which point partial neutralization of the alkali-impregnated final filter affects the results). The presence of acid gases other than hydrogen fluoride will also reduce the upper limit of detection.
The resultant determination for a sampling period greater than two days may indicate elevated particulate fluoride levels with depleted gaseous fluoride levels but the total value will still be valid (see Clause 7.2).
NOTES:
1 This method is subject to certain interferences. Particulate metallic salts, such as those of aluminium, iron, magnesium and calcium, or rare-earth elements, collected on the acidimpregnated prefilter may retain some or all of the gaseous fluorides. The presence of fluorophosphates or fluoroborates could contribute varying amounts of fluorides as they dissociate unpredictably.
Acidic aerosols or gases may neutralize or acidify the
alkali-impregnated final filter thus preventing the quantitative
collection of acidic gaseous fluorides. If this potential
interference is present, the decreased alkalinity of the sample
extract may provide relevant information. Although a
citric-acid-impregna
2 The addition of total ionic strength adjustment buffer (TISAB) eliminates a number of possible interferences. The effects of metallic ions are minimized by complexing agents. Variation in ionic strength is overcome by maintaining a high concentration of sodium chloride in the TISAB solution. The TISAB solution is buffered to pH 5 to 6 to avoid negative interference from the formation of complexes containing fluoride ions and positive interference from hydroxide ions being measured.