ASTM International - ASTM E2967-15
Standard Test Method for Assessing the Ability of Pre-wetted Towelettes to Remove and Transfer Bacterial Contamination on Hard, Non-Porous Environmental Surfaces Using the Wiperator
| Organization: | ASTM International |
| Publication Date: | 1 May 2015 |
| Status: | active |
| Page Count: | 10 |
| ICS Code (Other standards related to microbiology): | 07.100.99 |
| ICS Code (Chemicals for industrial and domestic disinfection purposes): | 71.100.35 |
significance And Use:
5.1 Microbial decontamination of environmental surfaces by wiping is subject to many variables (4), and failure to standardize them properly during testing of towelettes may give... View More
scope:
1.1 This standard is designed for use with a mechanized device (the Wiperator; Appendix X1) to test pre-wetted towelettes.
1.2 Two species of vegetative bacteria, one Gram-positive coccus (Staphylococcus aureus) and one Gram-negative bacillus (Acinetobacter baumannii), representing important nosocomial pathogens, are used to separately contaminate disks of magnetized and brushed stainless steel in order to test the towelettes for their relative ability to:
1.2.1 Decontaminate non-porous environmental surfaces experimentally-conta
1.2.2 Transfer any acquired bacterial contamination on the towelettes to clean surfaces.
1.3 This test method is not meant for use with towelettes for decontamination of skin.
1.4 The values stated in SI units are to be regarded as the standard. The values given in parentheses, if any, are for information only.
1.5 This test method should be performed by persons with training in microbiology in facilities designed and equipped for work with infectious agents at the appropriate biosafety level.
1.6 It is the responsibility of the investigator to determine whether Good Laboratory Practice (GLP) regulations are required and to follow them where appropriate.
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.