AAMI EC11
Diagnostic electrocardiographic devices
| Organization: | AAMI |
| Publication Date: | 1 January 1991 |
| Status: | inactive |
| Page Count: | 60 |
scope:
This standard establishes minimum safety and performance
requirements for electrocardiographic
NOTE-The safety and performance criteria defined in this standard are intended principally for use in design qualification or evaluation by the manufacturer.
The referee test methods of section 4 are intended to provide means by which conformance with the standard can be established unambiguously. These tests are not intended for use in verifying the performance of individual devices, either for purposes of quality assurance inspections by the manufacturer or for purposes of routine inhospital inspections. Also, referee tests, by definition, allow for the use of alternative methods for design qualification, provided that the equivalence of the methods can be established in terms of comparability of test results with those of the referee methods.
Inclusions
Included within the scope of this standard are the following devices:
a) direct-writing electrocardiographs;
b) electrocardiographs used in other medical devices (e.g., patient monitors, defibrillators, stress testing devices), when such devices are intended for use in obtaining diagnostic ECG signatures;
c) electrocardiographs having a display that is remote from the patient (via cable, telephone, telemetry, or storage media), when such devices are intended for use in obtaining ECG signatures. These devices are subject to the functional performance requirements at the system output-input levels2).
Exclusions
Not included within the scope of this standard are:
a) devices that collect ECG data from locations other than the external surface of the body;
b) devices for interpretation and pattern recognition (e.g., QRS detectors, alarm circuits, rate meters, diagnostic algorithms);
c) fetal ECG monitors;
d) ambulatory monitoring electrocardiographic
e) diagnostic electrocardiographic
f) vectorcardiographs, that is, the display of loops;
g) electrocardiographic
h) cardiac monitors, with or without heart rate meters and alarms, that are intended primarily for detecting cardiac rhythm. (These devices are covered by the ANSI/AAMI standard Cardiac monitors, heart rate meters and alarms.)
NOTE-Devices that provide selection between diagnostic and
monitoring functions must meet the requirements of the appropriate
standard-the standard for diagnostic electrocardiographic
1)Electrodes and associated cables are subject to the
provisions of this standard in as much as they are required for
conducting most of the performance and safety tests.
Manufacturer-specifi
2)It is recognized that the quality of telephone line or other means of transmission cannot be controlled by the electrocardiograph manufacturer. If the ECG equipment is intended for use with telephone transmission, however, the standard requires that the manufacturer disclose the minimum performance characteristics of the remote transmission system (e.g., allowable error, noise level, frequency response) necessary to ensure that the total system meets the requirements of the standard.
Document History