NEN - NPR-IEC/TS 62558
Ultrasonics - Real-time pulse-echo scanners - Phantom with cylindrical, artificial cysts in tissue-mimicking material and method for evaluation and periodic testing of 3D-distributions of void-detectability ratio (VDR)
| Organization: | NEN |
| Publication Date: | 1 April 2011 |
| Status: | active |
| Page Count: | 56 |
| ICS Code (Electroacoustics): | 17.140.50 |
scope:
This technical specification specifies essential characteristics of a phantom and method for the measurement of void-detectability ratio for medical ultrasound systems and related transducers. It is restricted to the aspect of long-term reproducibility of testing results. This technical specification establishes: - important characteristics and requirements for a TMM 3D artificial cyst phantom using anechoic voids; - a design example of a 3D artificial cyst phantom, the necessary test equipment and use of relevant computer software algorithms. This technical specification is currently applicable for linear array transducers. A uniformity test prior to void-detectability ratio (VDR) measurement is recommended.