UNLIMITED FREE
ACCESS
TO THE WORLD'S BEST IDEAS

SUBMIT
Already a GlobalSpec user? Log in.

This is embarrasing...

An error occurred while processing the form. Please try again in a few minutes.

Customize Your GlobalSpec Experience

Finish!
Privacy Policy

This is embarrasing...

An error occurred while processing the form. Please try again in a few minutes.

NASA-LLIS-0580

Lessons Learned - Adherence to Manufacturer's Recommendations and to Established Procedures for Hazardous Operations (Plastic Cryogenic Tissue Storage Vial Burst Hazard)

active, Most Current
Organization: NASA
Publication Date: 12 January 1998
Status: active
Page Count: 2
scope:

Description of Driving Event:

A small plastic vial stored in liquid nitrogen burst while laboratory workers were inventorying and transferring vials from one dewar to another. During the transfer a vial burst, causing superficial cuts on the hands of two laboratory workers.

The vial manufacturer had updated catalog and shipping documents to alert customers that the vials were not recommended for storage in the liquid phase of cyrogenically stored nitrogen. Instead, the manufacturer recommends that the vials be stored only in the gaseous phase of cryogenic nitrogen. The seal on the vials is not adequate to protect against seepage of liquid into the vial which could then burst following removal from the liquid unless adequate precautions were taken to allow the liquid to evaporate and escape slowly.

Workers were not wearing the selected PPE at the time of the mishap. The PPE had been removed because it was too bulky to permit manipulation of the small vials. This directly exposed the personnel to low temperature hazards as well as the burst hazard from the cryogen stored vials.

Document History

NASA-LLIS-0580
January 12, 1998
Lessons Learned - Adherence to Manufacturer's Recommendations and to Established Procedures for Hazardous Operations (Plastic Cryogenic Tissue Storage Vial Burst Hazard)
Description of Driving Event: A small plastic vial stored in liquid nitrogen burst while laboratory workers were inventorying and transferring vials from one dewar to another. During the transfer a...
Advertisement