DSF/PREN IEC 62220-2
Medical electrical equipment – Characteristics of digital X-ray imaging devices – Part 2: Determination of dual-energy subtraction efficiency – Detectors used for dual-energy radiographic imaging
| Organization: | DS |
| Status: | pending |
| Page Count: | 23 |
| ICS Code (Radiographic equipment): | 11.040.50 |
scope:
This document describes the performance metrics associated with DUAL-ENERGY IMAGING capable DIGITAL X-RAY IMAGING DEVICES meant for medical applications and specifies the methods for their determination. These metrics can be used to analyze TISSUE-SUBTRACTED IMAGES and to evaluate dose performance, noise characteristics, and tissue-subtraction efficacy of DIGITAL X-RAY IMAGING DEVICES. The described methods indicate the procedures to obtain MULTI-SPECTRAL PRIMARY DATA and to compute their derived TISSUE-SUBTRACTED IMAGES. The intended users of this part of IEC 62220 are MANUFACTURERS and well-equipped test laboratories. This document is restricted to DIGITAL X-RAY IMAGING DEVICES that are used for single or multiple exposure dual-energy radiographic imaging based on, for example, CR systems, direct and indirect flat panel-detector based systems. This document excludes and is not applicable to: - DIGITAL X-RAY IMAGING DEVICES intended to be used in mammography or in dental RADIOGRAPHY; - slot scanning DIGITAL X-RAY IMAGING DEVICES; - COMPUTED TOMOGRAPHY or CONE-BEAM COMPUTED TOMOGRAPHY; - photon-energy discriminating devices such as photon counting X-RAY IMAGING DEVICES; - devices for dynamic imaging (where series of images are acquired, as in fluoroscopy or cardiac imaging). - DIGITAL X-RAY IMAGING DEVICES intended to be used with RADIOTHERAPY beams