CEN - PREN 14626

Ambient air - Standard method for the measurement of the concentration of carbon monoxide by non-dispersive infrared spectroscopy

pending
Organization: CEN
Publication Date: 1 October 2022
Status: pending
Page Count: 101
ICS Code (Ambient atmospheres): 13.040.20
scope:

This document specifies a continuous measurement method for the determination of the concentration of carbon monoxide present in ambient air based on the non-dispersive infrared spectroscopic measuring principle. This document describes the performance characteristics and sets the relevant minimum criteria required to select an appropriate non-dispersive infrared spectroscopic analyser by means of type testing. It also includes the evaluation of the suitability of an analyser for use in a specific fixed site so as to meet the data quality requirements and requirements during sampling, calibration and quality assurance for use.

NOTE 1 Additional information is specified in Annex I of Directive 2008/50/EC [1]

The method is applicable to the determination of the mass concentration of carbon monoxide present in ambient air up to 100 mg/m3 of carbon monoxide. This concentration range represents the certification range for type testing.

NOTE 2 Other ranges can be used depending on the levels present in ambient air.

NOTE 3 When the standard is used for other purposes than for measurements required by Directive 2008/50/EC, the ranges and uncertainty requirements might not apply.

The method covers the determination of ambient air concentrations of carbon monoxide in locations classified as rural areas, urban-background areas, and for sampling points influenced by traffic or industrial sources.

The results are expressed in mg/m3 (at 20 °C and 101,3 kPa).

NOTE 4 100 mg/m3 of CO corresponds to 86 μmol/mol of CO.

This document contains information for different groups of users.

Clause 5 to Clause 7 and Annex B, Annex C and Annex D contain general information about the principles of carbon monoxide measurement by non-dispersive infrared spectroscopic analyser and sampling equipment.

Clause 8 and Annex E are specifically directed towards test houses and laboratories that perform type testing of carbon monoxide analysers. These sections contain information about:

- type testing conditions, test procedures and test requirements;

- analyser performance requirements;

- evaluation of the type testing results;

- evaluation of the uncertainty of the measurement results of the carbon monoxide analyser based on the type testing results.

Clause 9 to Clause 11 and Annex F are directed towards monitoring networks performing the practical measurements of carbon monoxide in ambient air. These sections contain information about:

- initial installation of the analyser in the monitoring network and acceptance testing;

- ongoing quality assurance/quality control;

- calculation and reporting of measurement results;

- evaluation of the uncertainty of measurement results under practical monitoring conditions.

The present document represents an evolution of earlier editions (EN 14626:2005 and EN 14626:2012). When equipment is procured it complies fully with the present document.

NOTE 5 Type testing performed prior to the publication of this document for the purpose of demonstrating equivalence are still valid.

NOTE 6 Analysers type tested prior to the publication of this document remain valid for use for regulated monitoring purposes.

Document History

PREN 14626
October 1, 2022
Ambient air - Standard method for the measurement of the concentration of carbon monoxide by non-dispersive infrared spectroscopy
This document specifies a continuous measurement method for the determination of the concentration of carbon monoxide present in ambient air based on the non-dispersive infrared spectroscopic...
August 1, 2012
Ambient air - Standard method for the measurement of the concentration of carbon monoxide by non-dispersive infrared spectroscopy
This European Standard specifies a continuous measurement method for the determination of the concentration of carbon monoxide present in ambient air based on the non-dispersive infrared...
March 1, 2005
Ambient air quality - Standard method for the measurement of the concentration of carbon monoxide by nondispersive infrared spectroscopy
A description is not available for this item.

References

Advertisement