E2911

Standard Guide for Relative Intensity Correction of Raman Spectrometers

active, Most Current
Publication Date: 1 April 2023
Status: active
Page Count: 12
scope:

This guide is designed to enable the user to correct a Raman spectrometer for its relative spectral-intensity response function using NIST Standard Reference Materials2 in the 224X series (currently SRMs 2241, 2242, 2243, 2244, 2245, 2246), or a calibrated irradiance source. This relative intensity correction procedure will enable the intercomparison of Raman spectra acquired from differing instruments, excitation wavelengths, and laboratories.

The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

Because of the significant dangers associated with the use of lasers, ANSI Z136.1 or suitable regional standards should be followed in conjunction with this practice.

This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

2 Trademark of and available from NIST Office of Reference Materials, 100 Bureau Drive, Stop 2300, Gaithersburg, MD 20899-2300.

Document History

E2911
April 1, 2023
Standard Guide for Relative Intensity Correction of Raman Spectrometers
This guide is designed to enable the user to correct a Raman spectrometer for its relative spectral-intensity response function using NIST Standard Reference Materials2 in the 224X series (currently...
July 1, 2013
Standard Guide for Relative Intensity Correction of Raman Spectrometers
This guide is designed to enable the user to correct a Raman spectrometer for its relative spectral-intensity response function using NIST Standard Reference Materials2 in the 224X series (currently...