ASTM International - ASTM F3018-25
Standard Guide for Assessment of Hard-on-Hard Articulation in Total Hip Joint Replacement and Resurfacing Hip Joint Replacement
| Organization: | ASTM International |
| Publication Date: | 1 July 2025 |
| Status: | active |
| Page Count: | 11 |
| ICS Code (Implants for surgery, prosthetics and orthotics): | 11.040.40 |
significance And Use:
4.1 This document provides guidance for a range of assessments and evaluations to aid in preclinical research and device development of hard-on-hard total and resurfacing hip joint replacements... View More
scope:
1.1 This guide covers materials and design recommendations and general test methods for the mechanical and preclinical assessment of implantable devices with hard-on-hard articulations intended to replace a hip joint. The provided guidance is intended to encompass both total hip joint replacements as well as resurfacing hip joint replacement (RS). There has been long-term clinical experience with metal-on-metal articulating components manufactured from cobalt-28chromium-6m
1.2 This guide applies to the acetabular and femoral articulating components of hard-on-hard hip joint replacements. Acetabular components can be monobloc, or a modular component with a separate cup and liner. As stated above, articulating components have been made from Co-28Cr-6Mo for a metal-on-metal bearing; and alumina-based ceramics for a ceramic-on-ceramic bearing. Cups of modular acetabular components have to date been made from Ti-6Al-4V (as specified in ASTM F136 or F1472, or ISO 5832-3) or Co-28Cr-6Mo. The shell is considered part of the acetabular component. Acetabular components may have an external coating and/or porous structure intended for uncemented, press-fit or biological fixation, or for use with bone cement.
1.3 This guide is a summary of available specifications, test methods, practices, and guides from published standards or the scientific literature. Their clinical relevance is unproven. Most of the methods do not have an established precision and bias; therefore, their repeatability and reproducibility have not been established. As the clinical relevance of these methods have not been established, consequently, most do not have performance requirements. This document does not require that all the listed methodologies are always necessary to evaluate these implant systems provided justification for not using each unused method is provided. This document does not intend to prevent the use of new methodologies as they are developed.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Document History