ASTM International - ASTM ISO/ASTM51401-03

Standard Practice for Use of a Dichromate Dosimetry System

inactive
Buy Now
Organization: ASTM International
Publication Date: 27 February 2003
Status: inactive
Page Count: 6
ICS Code (Radiation measurements): 17.240
significance And Use:

4.1 The dichromate system provides a reliable means for measuring absorbed dose in water. It is based on a process of reduction of dichromate ions to chromic ions in acidic aqueous solution by... View More

scope:

1.1 This practice covers the preparation, testing, and procedure for using the acidic aqueous silver dichromate dosimetry system to measure absorbed dose in water when exposed to ionizing radiation. The system consists of a dosimeter and appropriate analytical instrumentation. For simplicity, the system will be referred to as the dichromate system. It is classified as a reference standard dosimetry system (see ISO/ASTM Guide 51261).

1.2 This practice describes the spectrophotometric analysis procedures for the dichromate system.

1.3 This practice applies only to γ-rays, x-rays/bremsstrahlun, and high energy electrons.

1.4 This practice applies provided the following conditions are satisfied:

1.4.1 The absorbed dose range is from 2 X 10 3 to 5 X 104 Gy.

1.4.2 The absorbed dose rate does not exceed 600 Gy/pulse (12.5 pulses per second), or does not exceed an equivalent dose rate of 7.5 kGy/s from continuous sources (1).

1.4.3 For radionuclide gamma-ray sources, the initial photon energy shall be greater than 0.6 MeV. For bremsstrahlung photons, the initial energy of the electrons used to produce the bremsstrahlung photons shall be equal to or greater than 2 MeV. For electron beams, the initial electron energy shall be greater than 8 MeV.

Note 1-The lower energy limits given are appropriate for a cylindrical dosimeter ampoule of 12 mm diameter. Corrections for displacement effects and dose gradient across the ampoule may be required for electron beams (2). The dichromate system may be used at lower energies by employing thinner (in the beam direction) dosimeter containers (see ICRU Report 35).

1.4.4 The irradiation temperature of the dosimeter shall be above 0°C and should be below 80°C.

Note 2-The temperature coefficient of dosimeter response is known only in the range of 5 to 50°C (see 4.3). Use outside this range requires determination of the temperature coefficient.

1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Specific precautionary statements are given in 8.3.

Document History

November 19, 2013
Standard Practice for Use of a Dichromate Dosimetry System
1.1 This practice covers the preparation, testing, and procedure for using the acidic aqueous silver dichromate dosimetry system to measure absorbed dose to water when exposed to ionizing radiation....
ASTM ISO/ASTM51401-03
February 27, 2003
Standard Practice for Use of a Dichromate Dosimetry System
1.1 This practice covers the preparation, testing, and procedure for using the acidic aqueous silver dichromate dosimetry system to measure absorbed dose in water when exposed to ionizing radiation....
January 22, 2002
Standard Practice for Use of a Dichromate Dosimetry System
1.1 This practice covers the preparation, testing, and procedure for using the acidic aqueous silver dichromate dosimetry system to measure absorbed dose in water when exposed to ionizing radiation....